ANALISIS KUALITAS DATA KELUARAN HARIAN ELEKTRON AKSELERATOR LINIER

Heru Santoso, Muhammad Hamdi, Saktioto Saktioto

Abstract


Linac radiotherapy measurentoften experience instability. One of the irradiation errors with Linac can occur because the radiation beam that comes out is not expected properly. Determination of the correction factor and linearity is important to analyze the charge output of the electron energy emitted by Linac's modality to see the stability of the emitted charge. This study uses Electron Linear Accelerator Electron beam daily output data with a 10 cm × 10 cm applicator, 100 cm SSD, 0.125 cc cylindrical ionization detector PTW type 31010 Semiflek and uses a slab solid phantom with energy variations of 6 MeV, 8 MeV, 10 MeV, 12 MeV and 15 MeV. The measurement results are calculated using the IAEA TRS 398 protocol. The result of linearity correction factor of 96.87% which shows the stability value of the electron beam load output is very good, and the results of the correction factor show the enumeration values of each energy 6 MeV, 8 MeV, 10 MeV, 12 MeV and 15 MeV namely 0.030342129 nC, 0.03034 nC, 0.03034 nC, 0.03034 nC and 0.03034 nC values respectively still within the tolerance range of measurement ± 1. The correction factor that has been obtained is used as a parameter in calculating the absorbance dose to the maximum depth.


Keywords


Linac; Electron; Correction factor; TRS 398

Full Text:

PDF (INDONESIA)

References


1. Yeslem, A. E. M. O. M., Cheibetta, M. O., Ghassoun, J., Hakam, O. K., Semghouli, S., & Choukri, A. (2017, October). Quality control results for linear accelerator at Oncology Center in Nouakchott. Proceedings of the Mediterranean Symposium on Smart City Applications, 759–765.

2. Pratiwi, U. (2010). Aplikasi analisis citra detail phantom dengan metode konversi data digital ke data matrik untuk meningkatkan kontras citra menggunkan Film Imaging Plate. Skripsi, Fisika, Universitas Sebelas Maret.

3. Suharni, S., Kusminarto, K., & Anggraita, P. (2013). Analisis Hasil pengukuran Percentage Depth Dose (PDD) berkas elektron LINAC Elekta RSUP Dr. Sardjito. Prosiding Pertemuan dan Presentasi Ilmiah Teknologi Akselerator dan Aplikasinya, ISSN 1411-1349.

4. Nazaroh, N. Sunaryati, S. I., & Rajagukguk, N. (2006). Penerapan untuk jaminan kualitas radioterapi pusat teknologi keselamatan dan metrologi radiasi Badan Tenaga Nuklir Nasional. Seminar Keselamatan Nuklir.

5. IAEA. (2000). Technical Report Series No. 398 Absorbed Dose Determination in External Beam Radiotherapy. Vienna.

6. Liauw, S. L., Connell, P. P., & Weichselbaum, R. R. (2013). New paradigms and future challenges in radiation oncology: an update of biological targets and technology. Science translational medicine, 5(173).

7. Podgorsak, E. B. (2005). Radiation Oncology Physics: A Handbook for Teachers and Students. Vienna: IAEA.

8. Malicki, J. (2015). Medical physics in radiotherapy: The importance of preserving clinical responsibilities and expanding the profession's role in research, education, and quality control. Reports of Practical Oncology & Radiotherapy, 20(3), 161–169.

9. Rödel, C., Trojan, J., Bechstein, W. O., & Woeste, G. (2012). Neoadjuvant short-or long-term radio (chemo) therapy for rectal cancer: how and who should be treated?.Digestive Diseases, 30(2), 102–108.

10. Andreo, P. (2000). Absobed Dose Determination in External Beam Radioterapy. Vienna: IAIE.

11. Rusli, M. (2017). Uji Keselmatan Paparan Radiasi Dental Sinar-X di Radiologi Atro Muhammadiyah Makasar. Skripsi Fisika, Universitas Hasanudin.

12. Fauziah, A. S. & Dwijananti, P. (2013). Pengaruh Radiasi Sinar-X terhadap Mortilitas Sperma pada Tikus Mencit (Mus Musculus). Unnes Physics Journal, 2(2), 1–5.

13. Vadila, M. (2018). Analisis Keluaran Berkas Radiasi Pesawat Terapi Linac Tipe Varian CX 6264 di RS Unand. Skripsi Fisika, Universitas Andalas.

14. Güngör, G., Aydın, G., Mustafayev, T. Z., & Özyar, E. (2019). Output factors of ionization chambers and solid state detectors for mobile intraoperative radiotherapy (IORT) accelerator electron beams. Journal of applied clinical medical physics, 20(2), 13–23.




DOI: http://dx.doi.org/10.31258/jkfi.17.3.166-171

Refbacks

  • There are currently no refbacks.


Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

Indexing by:

  

 

Image